ANALISIS NILAI COMPUTED TOMOGRAPHY DOSE INDEX (CTDI) PADA PHANTOM KEPALA DI INSTALASI RADIOLOGI RUMAH SAKIT UMUM DAERAH SAYANG RAKYAT MAKASSAR
DOI:
https://doi.org/10.22437/jop.v10i1.37149Keywords:
CTDI;, CT Scan;, Dose Profiler;, Head Phantom;Abstract
This research aims to determine the CTDI value on a CT Scan aircraft using Dose Profiler and to determine the comparison of the CTDI (Computed Tomography Dose Index) value on the surface of each measurement position. The measurement method was carried out by measuring the CT Scan radiation dose with the CTDI value on the phantom using a RaySafe multimeter at a voltage of 130 kV, using a thickness of 10 mm, a time current accuracy of 100 mAs, 200 mAs and 300 mAs. The research results show the dose received on the surface of each type of phantom position with different time current accuracy respectively for a time current of 100 mAs at the center (9.139 mGy), type 1 (9.606 mGy), type 2 (9.170) type 3 (9.221) , and type 4 (9,210). A time current of 200 mAs is found at the center (9,110 mGy), edge 1 (9,578 mGy), edge 2 (9,546 mGy), edge 3 (9,224 mGy), and edge 4 (9,200 mGy). A time current of 300 mAs is found at the center position (18.39 mGy), edge 1 (19.33 mGy), edge 2 (19.18 mGy), edge 3 (18.48 mGy), and 4 (18.51 mGy) . Time-current variations affect the radiation dose that can be received by the phantom and have a linear influence on the CTDIvol value.
Downloads
References
Ahadiyah, N. N., Setiawati, E., & Arifin, Z. 2020. Pengukuran Computed Tomography Dose Index (CTDI) pada Fantom Kepala Dengan Menggunakan CT Dose Profiler. BERKALA FISIKA: Jurnal Fisika Teori, Eksperimen dan Fisika Aplikasi. 23(4). 118-125.
Al-Sharify, Z. T., Al-Sharify, T. A., Al-Sharify, N. T., & Yahya Naser, H. 2020. A Critical Review On Medical Imaging Techniques (CT and PET Scans) In The Medical Field. IOP Conference Series: Materials Science and Engineering, 870(1), 012043. https://doi.org/10.1088/1757-899X/870/1/012043
Amelia, C. 2014. Analisa CTDI pada Permukaan dan Pusat Phantom Menggunakan CT Dose Profiler. Youngster Physics Journal. 3(4). 329-334.
Bonilha, H. S., Wilmskoetter, J., Tipnis, S., Horn, J., Martin-Harris, B., & Huda, W. 2019. Relationships Between Radiation Exposure Dose, Time, and Projection In Videofluoroscopic Swallowing Studies. American journal of speech-language pathology, 28(3), 1053–1059. Https://doi.org/10.1044/2019_ajslp-18-0271
Chityala, R. N., Hoffmann, K. R., Bednarek, D. R., & Rudin, S. 2004. Region Of Interest (ROI) Computed Tomography (M. J. Yaffe & M. J. Flynn, Eds.; p. 534). https://doi.org/10.1117/12.534568
Dabukke, H. 2017. Pengaruh Perubahan Tegangan Terhadap Kontras Resolusi Pada CT Scan. Jurnal Mutiara Elektromedik. 1(1).
Harmayeni, H., Milvita, D., & Sandy, K. Y. P. 2019. Analisis Nilai CTDI di Udara dengan Variasi Faktor Eksposi dan Tebal Slice pada Pesawat CT-Scan Merek GE Optima 660. Jurnal Fisika Unand, 8(1), 52–56. https://doi.org/10.25077/jfu.8.1.52-56.2019
Hidayatullah, R. 2017. Dampak Tingkat Radiasi Pada Tubuh Manusia. Jurnal Universitas Sari Mutiara Indonesia. 1(1).
Hussain, S., Mubeen, I., Ullah, N., Shah, S. S. U. D., Khan, B. A., Zahoor, M., Ullah, R., Khan, F. A., & Sultan, M. A. 2022. Modern Diagnostic Imaging Technique Applications and Risk Factors in the Medical Field: A Review. BioMed Research International, 2022, 1–19. https://doi.org/10.1155/2022/5164970
Labania, H. M., P. Rindayani, Kasman, Abd. Rahman, & S. Ulum. 2021. Analisis Kontras Digital Radiography dengan Menggunakan ImageJ: (Analysis of Digital Radiography Contrast using ImageJ). Gravitasi, 20(1), 10–18. https://doi.org/10.22487/gravitasi.v20i1.15521
Milvita, D., Prasetio, H., & Yuliati, H. 2013. Perbandingan Dosis Radiasi di Udara Terhadap Dosis Radiasi Di Permukaan Phantom pada Pesawat CT-Scan. Jurnal Fisika Unand. 2(2). 114-119 https://doi.org/10.25077/jfu.2.2.%25p.2013
Irsal, M., Nurbaiti, N., Mukhtar, A. N., Gunawati, S., & Hidayat, W. 2021. Pengaruh Tegangan Tabung terhadap Kualitas Citra pada Pemeriksaan Computed Tomography Kepala menggunakan Iterative Reconstruction. Jurnal Teori dan Aplikasi Fisika, 9(1), 103–110. https://doi.org/10.23960/jtaf.v9i1.2715
Sari, D. A., & Setiawati, E. 2020. Analisis Nilai Computed Tomography Dose Index (CTDI) Phantom Kepala Menggunakan CT Dose Profiler Dengan Variasi Pitch. Berkala Fisika: Jurnal Fisika Teori, Eksperimen dan Aplikasi Fisika. 23(2).
Sookpeng, S., Cheebsumon, P., Pengpan, T., & Martin, C. 2016. Comparison Of Computed Tomography Dose Index In Polymethyl Methacrylate And Nylon Dosimetry Phantoms. Journal of Medical Physics, 41(1), 45. https://doi.org/10.4103/0971-6203.177287
Utari, E. L., & Listyalina, L. 2019. Analisis Reproduksibilitas Akurasi Tegangan di Pesawat Sinar-X Medis pada Bagian Thoraks Manusia. Prosiding Seminar Nasional Universitas Respati Yogyakarta. 1(2).
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2024 Ilham Ilham, Sri Zelviani, Jumardin Jumardin, Khaerul Bariah
This work is licensed under a Creative Commons Attribution 4.0 International License.